EasyManua.ls Logo

DENTSPLY Multimat2Sinter - 3 Technical Specifications; General; Electrical Connection; Operating Conditions

DENTSPLY Multimat2Sinter
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 11
3 Technical specifications
3.1 General
Parameter
Value
Unit
External dimensions (W × H × D)
385 × 780 × 500
mm
Heating chamber volume
2 bowls 120 mm × 35 mm
Maximum temperature
1650
°C
Heating rate
selectable: 199
°C/min
Weight
58
kg
3.2 Electrical connection
Electrical
Parameter
Value
Unit
Nominal supply voltage
220-240
V AC
Frequency
50
Hz
Permissible deviation
from the nominal voltage
± 10
%
Maximum Power
consumption
2000
W
Overvoltage category
II (pursuant to IEC 60364-4-443)
Protection against
overvoltage
Provide suitable surge filters
Fuse rating (on site)
Connect to a dedicated circuit protected by a
T10 AH 250 V fuse
3.3 Operating conditions
Environment
Parameter
Value
Unit
Temperature range
540
°C
Maximum relative
humidity
up to 31°C: 80
%
up to 40°C: 50
%
non-condensing
Installation
Indoors in dry rooms only
Height above sea level
maximum 2000
m
Pollution severity
Tested for pollution severity Pollution degree II