3300-0192-0M Rev D 01/17 18
Setup Menu Page 3
MENU PAGE 3
Decimal1: 1
Decimal2: 1
Motion Range: 3
Filter Mode: 0
<<< Exit >>>
Decimal1:
Touching this key will open the decimal entry screen. This will allow you to set the decimal
point precision for the first weight range. Using the numeric keys, enter a new value and
then press the Save key to save it.
Allowable values for decimal point precision are: 0 – 3.
Decimal2:
Touching this key will open the decimal entry screen. This will allow you to set the decimal
point precision for the second weight range if enabled (see Capacity2). Using the numeric
keys, enter a new value and then press the Save key to save it.
Allowable values for decimal point precision are: 0 – 3.
Motion Range:
Touching this key will open the motion range entry screen. This will allow you to set the
number of scale divisions of movement that will be allowed for stable. Using the numeric
keys, enter a new value and then press the Save key to save it.
Allowable values for motion are: 0 – 20.
Filter Mode:
Pressing this key will open the weight filtering screen. This will allow you to set the amount
of digital filtering being applied to the scale. Using the numeric keys, enter a new value and
then press the Save key to save it.
Allowable settings are: 0 = Off, 1 = Minimal, 2 = Moderate, and 3 = Maximum
Press the >>> key to advance to the Setup Menu Page 4.
Press the <<< key to return to the Setup Menu Page 3.
Press the Exit key to exit Setup and return to the Weight screen.