29
SE-DV53-1
EN
SpEcificationS
Problem Probable Cause Remedy
Condensation of
water is
collecting in the
hose causing a
gurgling noise.
1. Your humidification setting is
too high.
2. The room temperature
fluctuates from higher to lower
levels through the night.
Reduce the humidifier heater setting
and/or increase the temperature of
the room.
Air from flow
generator seems
to be too warm.
1. The air filter(s) is/are dirty.
2. The air inlet port is blocked.
3. The room temperature is too
high.
4. The SleepCube is located near
a heater.
1. Refer to the Cleaning section for
instructions on filter maintenance.
2. Unblock the air inlet.
3. Reduce the room temperature.
4. Locate the device away from the
heat source so the air being
drawn into the device is at room
temperature.
If the problem persists contact
your home care provider.
Nasal, sinus or
ear pain, possibly
with runny nose.
You may have a reaction to air
flow pressure.
Discontinue use and contact your
physician.
sPeCifiCaTions
Size . . . . . . . . . . . . . . . . . . . . . . . . 10.7 cm x 16.5 cm x 17.5 cm (4.2” H x 6.5” W x 6.9” D)
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.22 kg (2.7 lbs.)
Electrical Requirements AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240V~, 50/60 Hz
Electrical Requirements DC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption . . . . . . . . . . . . . . . . . . 65 watts max from AC power source
(flow generator only)
Pressure Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20 cmH2O
Operating Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . .5°C to 40°C (41°F to 104°F)
Operating Humidity Range. . . . . . . . . . . . . . . . . . . . . . . . . . 0% to 95% RH non-condensing
Operating Atmospheric Conditions . . . . . . . . . . . . . . . . . . . .Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range. . . . . . . . . . .-40°C to +70°C (-40°F to 158°F)
Storage & Transportation Humidity Range. . . . . . . . . . . . . . 0% to 95% RH non-condensing
Maximum Limited Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 cmH2O under normal use
Sound Pressure Level (tested per ISO 17510-1:2007) . . . . . . . . . . . . . . . . . . . . . . . . 26 dBA
Sound Power Level (tested per ISO 17510-1:2007) . . . . . . . . . . . . . . . . . . . . . . . . . . 38 dBA