30
SE-DV54-1
EN
SpEcificationS
Dynamic Short-Term Pressure Accuracy with and without
DV5HH Humidifier
(per ISO 17510-1:2007)
Waveform
Volume
[mL]
Breath Rate
[min-1]
Set Pressure
[cmH2O]
Pressure Accuracy
[Pk-Pk cmH2O]
Sinusoidal cycle with
I:E = 1:1
500 10, 15, 20
3.0 0.5
7.0 0.5
12.0 1.0
16.0 1.0
20.0 1.0
Equipment classification with respect to protection from electric shock . . . . . . . . . . .Class II
Degree of protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . Type BF Applied Part
Degree of protection against ingress of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX1
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
Mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Clinical Specifications for Digital Outputs
Mask Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 25.5 cmH2O ± 0.5
Patient Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -127 to +127 L/min ± 10 L/min
Leak Flow. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to +127 L/min ± 10 L/min
Estimated Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 1023 mL
Tidal Volume is an estimate. It is provided for trending purposes only.