29
SE-DV54-1
EN
SpEcificationS
Problem Probable Cause Remedy
Nasal, sinus or
ear pain, possibly
with runny nose.
You may have a reaction to air
flow pressure.
Discontinue use and contact your
physician.
sPeCifiCaTions
Size . . . . . . . . . . . . . . . . . . . . . . . . 10.7 cm x 16.5 cm x 17.5 cm (4.2” H x 6.5” W x 6.9” D)
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.22 kg (2.7 lbs.)
Electrical Requirements AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240V~, 50/60 Hz
Electrical Requirements DC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption . . . . . . . . . . . . . . . . . . 65 watts max from AC power source
(flow generator only)
Pressure Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20 cmH2O
Operating Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . .5°C to 40°C (41°F to 104°F)
Operating Humidity Range. . . . . . . . . . . . . . . . . . . . . . . . . . 0% to 95% RH non-condensing
Operating Atmospheric Conditions . . . . . . . . . . . . . . . . . . . .Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range. . . . . . . . . . .-40°C to +70°C (-40°F to 158°F)
Storage & Transportation Humidity Range. . . . . . . . . . . . . . 0% to 95% RH non-condensing
Maximum Limited Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 cmH2O under normal use
Sound Pressure Level (tested per ISO 17510-1:2007) . . . . . . . . . . . . . . . . . . . . . . . . 26 dBA
Sound Power Level (tested per ISO 17510-1:2007) . . . . . . . . . . . . . . . . . . . . . . . . . . 38 dBA
Filter Specifications
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Maximum Flowrate (per ISO 17510-1:2007)
Test Pressures
3.0 cmH20 7.0 cmH20 12.0 cmH20 16.0 cmH20 20.0 cmH20
Measured
pressure at the
patient connection
port (cmH20)
2.9 6.4 11.0 15.0 19.0
Average flow at
the patient
connection port
(l/min)
83.8 126.8 137.4 130.0 113.8
Long Term Static Pressure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +/- 0.5 cmH2O