33
SE-DV54-1
EN
guiDancE & manufacturEr’S
DEclaration
Guidance and Manufacturer’s Declaration – Emissions All Equipment and
Systems
This device is intended for use in the electromagnetic environment specified below. The
customer or user of this device should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level Electromagnetic Environment - Guidance
Power
Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m 3A/m
Power frequency magnetic fields should be
that of a typical location in a typical
commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
from
150 kHz to
80 MHz
V1 = 3
Vrms
Portable and mobile RF communications
equipment should be separated from the
device by no less than the recommended
separation distances calculated/listed
below:
D=(3.5/V1)√
P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
E1 = 3V/m
D=(3.5/E1)√
P
80 to 800 MHz
D=(7/E1)√
P
800 MHz to 2.5 GHz
Where P is the maximum power rating in
watts and D is the recommended
separation distance in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site
survey, should be less than the compliance
levels (V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance D in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.