1312
Scatter Results using Lead Shield
Horizontal Plane,
Significant Zone of Occupancy:
60 cm x 60 cm
Vertical Plane,
Significant Zone of Occupancy:
60 cm x 200 cm
2.3 Usage and Duty Cycle
As a safety feature, the NOMAD Pro 2 will not emit X-rays with insufficient voltage (low
battery).
The NOMAD Pro 2 is also designed to avoid damage from overheating. The minimum
duty cycle rating for maximum exposure (the relationship between duration and
frequency of exposures taken during a rolling 60 second period) is 1:60. This duty cycle is
programmed into your NOMAD Pro 2.
The NOMAD Pro 2 is a reusable device, but depending on usage patterns, handling, and
maintenance, service life may vary. If any of the conditions below are observed, the device
should be immediately removed from service. In this case, it may be possible for the
device to be factory repaired and returned to service for the remainder of its design life.
Contact Customer Service for details.
• Reduction in performance
• Cracks on the housing, collimator cone, and/or backscatter shield
• System dropped
• Subject to moisture (wetting, immersing, or soaking)