6.0 NOMAD Pro 2 Care and Upkeep
6.1 Alarms and Alerts
The visual and audible alarms signal a programmed action designed to prevent harm to
operators, patients, and/or the NOMAD Pro 2. The visual/audible alerts confirm normal
conditions or draw the operator’s attention to a required action.
NOTE: All audible signals except X-ray termination and completion may be turned down
or off in the Audible Sounds menu. See Section 4.10 Settings Menu.
Alerts & AlArms: VisuAl And Audible indicAtors
Function/resolution
VisuAl Audible
Enabling Alert None
This alert is initiated by a single pull
and release of the trigger, within
one second. If no further action
is taken, the device will shift into
the Ready state. If the trigger is
pulled again before the Ready
state is indicated, the second pull is
disregarded.
Ready Alert
Double ascending
tone
At the conclusion of the Enabling
X-Rays alert, the device will shift
into the Ready state. The Ready
state will last for 15 seconds, or
until one of the buttons on the user
interface panel is pressed, or until
the exposure is initiated by a second
pull of the trigger.