16
int= (INTERVAL SETTING)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are: 1, 2 or 5.
Unit= (WEIGHING UNIT)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are: 1, 2, 3 or 4.
1 = Pounds Only 2 = Kilograms Only
3 = Pounds/Kilograms 4 = Kilograms/Pounds
5 = Pounds & Ounces/Kilograms (if USA=1)
trA= (ZERO TRACKING RANGE)
Zero tracking range is a value in scale divisions that will be automatically zeroed off. Press
the ENTER/PRINT key to view the current setting. If the value displayed is acceptable,
press the ENTER/PRINT key again. Otherwise, press the or keys and the or keys
to scroll through and select a new value and then press the ENTER/PRINT key to save it
and proceed to the next prompt. Allowable values are 1 through 18 (1 to 9 divisions by 0.5
divisions). Select 0 (zero) to disable zero tracking.
UnS= [MOTION (UNSTABLE) RANGE]
The Motion (Unstable) Range is the number of divisions of change permitted before
the STABLE annunciator turns off.
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys
and the or keys to scroll through and select a new value and then press the
ENTER/PRINT key to save it and proceed to the next prompt. Allowable values are 0
through 99.
dPP= (DECIMAL POINT POSITION)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are: 0, 1, 2 or 3.
0 = XXXXXX 1= XXXXX.X 2 = XXXX.XX 3 = XXX.XXX
CAP= (SCALE CAPACITY)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys
and the or keys to scroll through and select a new value and then press the
ENTER/PRINT key to save it and proceed to the next prompt. Allowable values are: 1
through 99,999.
Setup & Calibration (cont.)