FLT= (DIGITAL FILTER LEVEL SELECTION)
Your indicator will arrive with factory filter settings of 1 = Minimal. Please check with
Tech Support before changing filter level, break range and sample rate.
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are:
0 = Minimal Filter 1 = Moderate Filter 2 = Heavy Filter 3 = Custom Filter
NOTE! If 3 = Custom Filter is selected, two additional prompts will be displayed.
F= (FILTER LEVEL)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are 1 (least amount of filtering)
to 99 (greatest amount of filtering).
BR= (BREAK RANGE)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are 1 to 99 which correspond to
the number of division changes to break out of filtering.
SR= (SAMPLE RATE)
Press the ENTER/PRINT key to view the current setting. If the value displayed is
acceptable, press the ENTER/PRINT key again. Otherwise, press the or keys to
scroll through and select a new value and then press the ENTER/PRINT key to save
it and proceed to the next prompt. Allowable values are a minimum of 1 sample per
second to a maximum of 16 samples per second in one sample per second intervals.
CAL= (CALIBRATION)
With the display showing CAL=, press the ENTER/PRINT key. The display will change
to show the current setting 0 (0=NO). If the scale has been previously calibrated and
you wish to skip calibration and proceed to the setGC (Set Gravity Constant) prompt,
press the ENTER/PRINT key and the previous calibration will be retained.
To begin calibration, press the key to select 1 (yes) and then press the ENTER/PRINT
key. After pressing the ENTER/PRINT key, the display will change to LOAd=.