18
LoAd= (LOAD CALIBRATION WEIGHT)
With the display showing LoAd=, perform the following steps:
1. Make certain the scale platform is empty and free of debris.
2. Place the desired amount of calibrated test weights on the scale platform. A minimum
of 50% of scale’s capacity is required. However 70% to 100% is recommended.
3. Press the ENTER/PRINT key.
4. If the value displayed is acceptable, press the ENTER/PRINT key again.
Otherwise, determine the exact amount of test weight placed on the scale
platform and then using the or keys and the or keys scroll through and
select the test weight amount.
5. Verify that the numbers selected are the same as the amount of the test weight
and then press the ENTER/PRINT key.
6. Starting at the left and proceeding right, a series of dashes will appear on the
display. The dashes will stay on the display momentarily, then disappear, after
which the display will proceed to the next prompt.
UnLoAd= (UNLOAD CALIBRATION WEIGHT)
After a moment, the display will change to UnLoad.
1. Remove the test weights from the scale platform and then press the ENTER/
PRINT key.
2. Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
calibration factor will be saved and the display will proceed to the next prompt.
IMPORTANT! During the time the dashes are appearing on the display, ensure
that the loaded (or empty) scale is stable.
SEtGC (SET GRAVITY CONSTANT)
This indicator is equipped with an acceleration of gravity function which means that
it can be calibrated in one location and then adjusted to match the acceleration of
gravity at the location where it will be used.
With the display showing setGC, press the ENTER/PRINT key to view the current
setting. If the value displayed is acceptable, press the ENTER/PRINT key again.
Otherwise, press the or keys to scroll through and select a new value and then
press the ENTER/PRINT key to save it and proceed to the next prompt. Allowable
values are: 0 or 1.
0 = Gravity Constant is NOT used 1 = Set Gravity Constant
NOTE! If you select 1 (Set Gravity Constant) the following additional prompts
will be displayed:
Setup & Calibration (cont.)