The bottom float of the brine sensor will stay in the up position in a normal condition. When the sensor is in an
alarm condition, the upper float will be in a position to cause the alarm condition or the fluid level has
decreased below the bottom float.
Cutaway View of Sensor Showing Internal Floats
Specifications
Primary Use: Measure the level of brine solution
Detects: Low Liquid, High Liquid
Operating Temperature: -40°C to +70°C (-40°F to 158°F)
D-10B Dimensions:
D-20B Dimensions:
Diameter: 5.8 cm (2.3 in.), Length: 28.2 cm (11.1 in.)
Diameter: 5.8 cm (2.3 in.), Length: 53.6 cm (21.1 in.)
Float Requirements: Low: 3.8 cm (1.5 in.), High: 27.9 cm (11 in.)
Cable:
Belden #88760 or Alpha #55371
3.6 m (12 feet) of gas & oil resistant cable to the inline ISIM
+ 1.3 m (4 feet) ISIM tail
Maximum Wiring Length*: 305 m (1,000 ft.) field wiring
Alarm Threshold Configuration: Fully Automatic
Diagnostic Reading on sensor setup:
3 to 4 (normal),
12 to 13 (bottom float in alarm - bottom and top floats in
the down position),
1 to 2 (upper float in alarm - top and bottom floats in the up
position)
Multi-Drop Restriction
Maximum points per channel = 12 points
924B and DMP Probes = 4 points
ISIM Smart Sensors = 1 point
M2050-PLUS MagLink LX Plus Installation Manual | Dover Fueling Solutions 83