178 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11.3.6.4 Vertical tab "Logbook"
Central O
2 supply On; Off Specifies which gas supplies are tested
during the automatic system test.
Central Air supply On; Off
Central N
2O sup-
ply
On; Off
O
2 cylinder
Air cylinder
N
2O cylinder
On; Off
Verify O
2 delivery On; Off Specifies whether the system test will
check that the O
2 supply is actually
delivering oxygen.
The Sample line is connected during
test parameter must also be set to On
for this.
Test breathing circuit
Test correct
assembly of
breathing hoses
and Y-piece
On; Off Specifies whether the automatic tests
will check that the breathing circuit is
correctly connected.
During this test, a check is made as to
whether the breathing gas can flow
from the inspiratory port via the Y-piece
to the expiratory port.
If Sample line is connected during test
is set to On, this test is performed fully
automatically. If Sample line is con-
nected during test is set to Off, addi-
tional manual checks are required.
Headline/
Parameter
Setting range
Description
A logbook entry with measured values is created Enables the creation of additional log-
book entries with measured values.
Every 1 min; 2 min; 5 min; 10 min; 15 min Generates periodic entries.
For all high-prior-
ity alarms
On; Off Generates entries in the event of an
alarm.
For all medium-
priority alarms
On; Off
Headline/
Parameter
Setting range
Description