Instructions for use BiliLux 81
Technical data
14.10.5 Recommended separation distances
The BiliLux is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or user of the BiliLux can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment and the BiliLux as
recommended, according to the maximum output power of the communications
equipment.
Maximum out-
put power
(Watts)
Separation (m)
150 kHz to 80-
MHz
Separation (m)
80 MHz to
800 MHz
Separation (m)
800 MHz to
2.5 GHz
0.01 0.1167 m
(0.3829 ft)
0.12 m
(0.39 ft)
0.23 m
(0.76 ft)
0.1 0.369 m
(1.211 ft)
0.38 m
(1.25 ft)
0.73 m
(2.4 ft)
11.167m
(3.829 ft)
1.2 m
(3.9 ft)
2.3 m
(7.6 ft)
10 3.69 m
(12.11 ft)
3.8 m
(12.5 ft)
7.3 m
(23.9 ft)
100 11.67 m
(38.29 ft)
12 m
(39 ft)
23 m
(76 ft)
D1.2P=
D1.2P=
D2.3P=

Table of Contents

Other manuals for Dräger BiliLux

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger BiliLux and is the answer not in the manual?

Dräger BiliLux Specifications

General IconGeneral
BrandDräger
ModelBiliLux
CategoryMedical Equipment
LanguageEnglish

Related product manuals