EasyManuals Logo

Dräger Fabius plus XL User Manual

Dräger Fabius plus XL
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Supplement Fabius plus XL SW 3.n 17
Operation
Safety functions of the ventilator
Behavior of Fabius if the user does not
take any action
The following safety information has changed:
WARNING
Risk of patient regaining consciousness or
risk of hypoxia
If the gas supply fails, further operation of the
anesthesia machine takes place with ambient
air. Oxygen and anesthetic gas are no longer
delivered. Due to the patient’s rebreathing,
the inspiratory concentration of anesthetic
gases and oxygen in the breathing gas falls.
Monitor the gas concentrations carefully and
use intravenous anesthetic agents if
necessary.

Other manuals for Dräger Fabius plus XL

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus XL and is the answer not in the manual?

Dräger Fabius plus XL Specifications

General IconGeneral
BrandDräger
ModelFabius plus XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals