Supplement Fabius plus XL SW 3.n 17
Operation
Safety functions of the ventilator
Behavior of Fabius if the user does not
take any action
The following safety information has changed:
WARNING
Risk of patient regaining consciousness or
risk of hypoxia
If the gas supply fails, further operation of the
anesthesia machine takes place with ambient
air. Oxygen and anesthetic gas are no longer
delivered. Due to the patient’s rebreathing,
the inspiratory concentration of anesthetic
gases and oxygen in the breathing gas falls.
Monitor the gas concentrations carefully and
use intravenous anesthetic agents if
necessary.