6 Device settings
Only trained and qualified personnel may change the
device settings.
For more details, refer to the technical manual.
6.1 Factory settings
Different settings can be selected to meet customer
requirements on order. The setting can be checked and
changed with the Dräger CC-Vision PC software.
The modified parameter settings must be checked on
the gas detector after transmission to ensure that the
values were transmitted correctly.
Parameters that cannot be viewed on the gas detector
must be exported using the Dräger CC-Vision PC
software after a change and checked.
The following factory settings are only available for the
X-am 8000:
Function Setting
Fresh air calibration with-
out password
On
Bump test without pass-
word
Off
Operation signal On
Switch off allowed On
Capture range
1)
On
LEL factor
2)
ch4 (methane)
H2 (hydrogen)
4.4 vol. % (corresponds to
100 %LEL)
4.0 vol % (corresponds to
100 %LEL)
STEL STEL feature – inactive;
average duration =
15 minutes
TWA TWA feature – inactive;
average duration =
8 hours
Alarm thresholds configu-
ration type
ATEX-compliant
Alarm A1 Acknowledgeable; not
latching, pre-alarm,
increasing measured
value (for O2 sensor,
decreasing measured
value as well)
Alarm A2 Not acknowledgeable;
latching, main alarm,
increasing measured
value (for O2 sensor,
decreasing measured
value as well)
Symbol for type of danger On
Expired bump test interval Warning
Expired calibration interval Channel warning
Power-save mode display Off
1) The set capture range can be read out on the gas detector
and activated or deactivated. The capture range is activated
in measuring mode ex-factory. The capture range is always
deactivated in calibration mode.
2) An LEL factor can be adapted to national regulations using
the Dräger CC-Vision PC software.
Function Setting
Bluetooth
®
(if installed)
Off
ToxicTwins (HCN) Off
No measured values in
the Vol% range
Off
Function Setting