6.3 Alarm settings (factory setting)
Definitions:
Pre-acknowledgement: If the alarm condition is acknowledged (by pressing the OK button), the acoustic alarm and
the vibration are switched off. The alarm is only completely reset (LED and display) once the alarm condition no
longer exists.
Acknowledgement: If an A1 alarm condition that no longer exists is acknowledged (by pressing the OK button), all
alarm elements are reset.
If the A2 and A1 alarms are configured so that they can be acknowledged, the pre-acknowledgement or
acknowledgement of the A2 alarm pre-acknowledges the A1 alarm, or completely acknowledges this alarm if the
alarm condition no longer exists.
Explanation of the symbols:
: Feature activated
: Pre-acknowledgement
Alarms/events Presentation
in the display
Latching
Acknowl-
edgeable
LEDs Horn Vibration
A1 (increasing)
A1 -
A2 (increasing)
A2 -
A1 (decreasing)
A1 -
A2 (decreasing)
A2 -
STEL
1)2)
1) The STEL alarm can be triggered with a delay of max. 1 minute.
2) After this alarm, the deployment of personnel is subject to the relevant national regulations.
STEL -
TWA
3)
3) A TWA alarm can only be reset by turning the gas detector off and on.
TWA -
Error
4)
4) Troubleshooting, see technical manual.
Battery pre-alarm
5)
5) The battery still lasts approx. 20 minutes after the battery pre-alarm triggers.
--
Battery main alarm
6)
6) The gas detector switches off automatically 20 s after a battery main alarm.
---
Device error
-
Channel error
- ---

Table of Contents

Other manuals for Dräger HFG 000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger HFG 000 Series and is the answer not in the manual?

Dräger HFG 000 Series Specifications

General IconGeneral
BrandDräger
ModelHFG 000 Series
CategoryGas Detectors
LanguageEnglish

Related product manuals