DRAFT 20 June 2005
Jaundice Meter Instructions for Use (MU01380) Page 2 - 13
SPECIFICATIONS
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter is intended for use in the electromagnetic envi-
ronments specified below. The customer or user of the unit should
ensure that the unit is used in such environments.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—
Guidance
Electrostatic
discharge
(ESD)—
IEC 61000-
4-2
± 6 kV con-
tact
± 8 kV air
± 6 kV con-
tact
± 8 kV air
The floors should be
wood, concrete, or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical
fast tran-
sient/burst
—IEC
61000-4-4
± 2 kV for
power sup-
ply lines
± 2 kV for
power sup-
ply lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge—IEC
61000-4-5
± 1 kV
differential
mode
± 1 kV
differential
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.