DRAFT 20 June 2005
Jaundice Meter Instructions for Use (MU01380) Page 2 - 15
SPECIFICATIONS
Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Jaundice Meter is intended for use in the electromagnetic environ-
ments specified below. The customer or user of the unit should ensure
that the unit is used in such environments.
Immunity
Test
IEC
60601
Test
Level
Compli
ance
Level
Electromagnetic Environment—
Guidance
Recommended Separation Distance
Con-
ducted
RF—IEC
61000-4-6
3 Vrms
150
kHz to
80
MHz
3 Vrms Portable and mobile RF communica-
tion equipment should be used no
closer to any part of the Jaundice
Meter, including cables, than the rec-
ommended separation distance calcu-
lated from the equation applicable to
the frequency of the transmitter.
Recommended Separation Distance
Radiated
RF—IEC
61000-4-3
3 V/m 3 V/m
80 MHz to 800 MHz
80
MHz
to 2.5
GHz
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended separa-
tion distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey
a
, should be less than
the compliance level in each frequency
range.
b
Interference may occur in the vicinity
of equipment marked with the follow-
ing symbol:
d 1.2 P=
d 2.3 P=