Care
Dismantling
Care
After
each
operational
use
of
the
ventilator,
dismantle
and
disinfect
the
ventilation
valve,
ventilation
hose,
volume
metering
parts
and
PEEP
valve.
The
ventilator
and
compressed
gas
hoses
must
also
be
cleaned/disinfected
if
severely
soiled.
Dismantling
Remove
Oxylog
1000
from
its
holder.
1
Unscrew
the
02
compressed
gas
hose
from
the
Oxylog
1000.
2
Remove
the
ventilation
hose
from
the
connector
on
the
side
of
the
ventilator.
To
remove
the
ventilation
hose,
always
grip
it
by
the
sleeve
and
not
the
spiral
stiffening
ridges.
Otherwise
the
spiral
ridges
may
be
torn,
e.g.
at
the
sleeve,
or
the
hose
may
be
ripped
out
of
the
sleeve.
Remove
the
ventilation
valve
from
the
ventilation
hose.
Pull
the
PEEP
valve
out
of
the
ventilation
valve,
or
remove
and
dismantle
the
volume
metering
parts.