Operation
Oxylog
1000
Device
Check
Operation
Oxylog
1000
Device
Check
The
device
check
must
be
carried
out
before
each
use.
Any
operation
of
the
device
requires
thorough
knowledge
of
the
Instructions
for
Use.
Type:
Serial
No.
Check
the
following
points
before
starting
up
the
device:
02
pressure
supply
connected
Cylinder
pressure
at
least
100
bar
or
ventilator
connected
to
the
central
O
2
supply
Ventilation
valve
and
ventilation
hose
connected
Testing
correct
operation
Fit
the
test
lung
to
the
ventilation
valve.
Set
the
device:
MV
Freq.
Paw
Alarm
Main
switch
Switch
approx.
10
L/min
approx.
20
bpm
approx.
55
mbar
I
(ON)
No
Air
Mix
Oxylog
1000
must
ventilate
the
test
lung.
Squeeze
the
test
lung,
so
that
the
airway
pressure
is
approx.
60
mbar.
The
»Paw
J
indicator
turns
red,
and
the
audible
alarm
is
sounded.
Remove
the
test
lung:
The
»Paw
«
indicator
turns
red,
and
the
audible
alarm
is
sounded.
In
the
event
of
deviations,
see
"Troubleshooting"
in
these
Instructions
for
use.
Device
check
completed
Name:
_
Date:
_
6