Care
Disinfecting/Cleaning
Dismantling
the
ventilation
valve
1
Turn
the
cover
about
45°
counter-clockwise
=
unlock,
and
remove
the
cover.
2
Remove
the
diaphragm.
Do
not
dismantle
any
further.
Do
not
allow
any
foreign
matter
to
enter
the
housing
of
the
ventilation
valve.
Protect
the
diaphragm
from
damage.
Disinfecting/Cleaning
Use
only
preparations
classified
as
»surface
disinfectants"
for
disinfecting.
For
material
compatibility,
we
recommend
preparations
based
on
-
aldehydes,
-
alcohols,
-
quaternary
ammonium
compounds.
Due
to
possible
damage
to
materials,
the
following
preparations
are
unsuitable:
-
compounds
containing
alkylamine
-
compounds
containing
phenol
-
halogen-releasing
compounds
-
strong
organic
acids
-
oxygen-releasing
compounds.
For
users
in
the
Federal
Republic
of
Germany,
we
recommend
the
use
of
disinfectants
listed
in
the
current
DGHM
list
(DGHM
=
German
Society
for
Hygiene
and
Microbiology).
The
DGHM
list
(published
by
mhp
Verlag
GmbH,
Wiesbaden)
also
specifies
the
active
basis
of
each
disinfectant.
For
countries
where
the
DGHM
list
is
unavailable,
we
recommend
products
based
on
the
above
active
bases.
12