16
Emergency Start Ups
Release the oxygen safety
control knob by pressing and
set to desired oxygen ow,
between 0 and 12 l/min.
The oxygen ow is channelled
through the vaporiser.
Manual ventilation is available
at any time.
Switch on Primus.
After approximately 35 seconds the checklist will appear.
Press the cancel test key (1).
You can cancel the self test up
to 10 times. After that, it is
mandatory to perform a self test.
Primus switches to standby
mode after completing the
minimal test (duration 10 seconds).
Close safety O
2
ow control.
Continue as in normal
operation.
1
Emergency Scenarios

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus Series and is the answer not in the manual?

Dräger Primus Series Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor Touchscreen
Power SupplyAC Mains (100-240V, 50/60Hz)
Backup BatteryMinimum 30 minutes (depending on configuration and usage)
Tidal Volume Range20 to 1500 mL
Respiratory Rate4 to 100 bpm
Inspiratory Time0.2-5.0 sec
PEEP0-20 mbar
Gas SupplyO2, N2O, Air
AlarmsApnea, Gas Supply
Ventilation ModesManual, Pressure Support, SIMV
O2 Flush35 to 45 L/min