Checking Compact Breathing
System
Ensure that the compact
breathing system (1) is connected
and the ventilator is locked in place.
Check status of CO
2
absorbent (2), breathing bag
and patient hoses.
In Addition …
Test the oxygen ush (3).
Check oxygen safety control (4).
Make sure vaporisers are in
position, locked and lled (5).
Check water trap level and the
sample gas line connection (6).
Check function of suction
system (7). Your suction system may
differ from the system shown.
Connect anaesthetic gas
scavenging system.
5
1 34
2
6
7
5
Preparation for Use

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus Series and is the answer not in the manual?

Dräger Primus Series Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor Touchscreen
Power SupplyAC Mains (100-240V, 50/60Hz)
Backup BatteryMinimum 30 minutes (depending on configuration and usage)
Tidal Volume Range20 to 1500 mL
Respiratory Rate4 to 100 bpm
Inspiratory Time0.2-5.0 sec
PEEP0-20 mbar
Gas SupplyO2, N2O, Air
AlarmsApnea, Gas Supply
Ventilation ModesManual, Pressure Support, SIMV
O2 Flush35 to 45 L/min