Standby Screen
Primus activates the standby mode after successful test completion,
and will display the standby screen.
From the standby screen, you can access the following functions:
Pre set carrier gas, O
2
concentration and fresh gas ow
Set age
Set ideal body weight
Restore site default settings
Set alarm limits
Self test results display
Perform the leak test
Delete trends
Default conguration
(password protected - held by Key Users and Dräger)
Access monitoring mode
Monitoring Mode
In monitoring mode, the Primus will measure and display gas measurements.
There is no fresh gas delivery or ventilation capability. If ventilation is required,
set the appropriate gas ow, select the appropriate ventilation mode and
press the rotary control knob to conrm.
9
Operation

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus Series and is the answer not in the manual?

Dräger Primus Series Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor Touchscreen
Power SupplyAC Mains (100-240V, 50/60Hz)
Backup BatteryMinimum 30 minutes (depending on configuration and usage)
Tidal Volume Range20 to 1500 mL
Respiratory Rate4 to 100 bpm
Inspiratory Time0.2-5.0 sec
PEEP0-20 mbar
Gas SupplyO2, N2O, Air
AlarmsApnea, Gas Supply
Ventilation ModesManual, Pressure Support, SIMV
O2 Flush35 to 45 L/min