Leak Test
The leak test should be carried out after the patient hoses, vaporisers or soda lime have
been replaced. In addition, it should be performed following the self test.
To perform the leak test:
In the standby screen, press the leak test soft key
Follow the on screen instructions
Push the rotary control knob to start the leak test
Self test results can be displayed, see page 7
Self Test Colour Result Action Example
Outcome
Green Test completed No action 100% functional
successfully required
Yellow Primus can be Faults must be Loudspeaker failure
used with reported or remedied
restrictions immediately after use
Red The device has If the fault is recognised O
2
failure
a malfunction immediately and can be
instantly corrected, the
test is repeated.
If operation is not
possible or not permitted
the test cannot be
repeated and the device
is taken out of use
There are three potential outcomes of the self test:
8
Start Up

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus Series and is the answer not in the manual?

Dräger Primus Series Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor Touchscreen
Power SupplyAC Mains (100-240V, 50/60Hz)
Backup BatteryMinimum 30 minutes (depending on configuration and usage)
Tidal Volume Range20 to 1500 mL
Respiratory Rate4 to 100 bpm
Inspiratory Time0.2-5.0 sec
PEEP0-20 mbar
Gas SupplyO2, N2O, Air
AlarmsApnea, Gas Supply
Ventilation ModesManual, Pressure Support, SIMV
O2 Flush35 to 45 L/min

Related product manuals