Self Test
Follow the self test instructions
on the screen.
Information about system leaks
and compliance will be displayed.
Primus will accept leaks of up to
150 mL/min. The most common
causes of leakage are:
Faulty breathing hoses or breathing bag
Sample gas line
Filter
Incorrectly connected absorber
Flow sensors
Breathing system
Vaporisers
Water trap
Oxygen sensor
Potential Test Results
All test results are colour coded and
are displayed next to each specic
test function. An advisory window
containing information on how to
remedy the problem accompanies
any error detected during the self test.