Start Up
Switch on Primus using the
ON/OFF switch (1).
Close Y-piece (place rmly
onto the corresponding cone
of the breathing system).
Set the APL valve to MAN
and 30 mbar (2). There are
two versions of the APL valve.
Your version may differ from
the one shown.
Close oxygen safety control
valve (3).
Conrm the checklist,
if all points are OK.
Primus will run internal
system tests.
Note: During this start up
procedure and up until the
checklist is conrmed, manual
ventilation is always possible. The
oxygen safety ow can be set to
0 -12 L/min and will be channelled
through the vaporiser.
6
2 3 1
Start Up