19
Ventilator Failure
In case of ventilator error, the
ventilator will stop and display the
error message “VENTILATOR
FAIL” on the screen.
Fresh gas ow and full
monitoring functionality are
not affected.
Operation automatically switches
to Man. Spont.; only manual
ventilation is possible.