Central
gas
supply
ZV
Gas
pressures:
1.
All
LEDs
light
up
green
if
pressure
values
are
available
and
are
between
2.7
and
6.9
bar.
The
LEDs
remain
dark
if
the
gas
pressure
is
<2.7
bar
or
if
the
gas
hose
is
not
connected.
Backup
gas
cylinders
2
Slowly
open
the
cylinder
valves.
LEDs
light
up
green
when
O2
pressure
is
over 20
bar
and
N20
pressure
is
over
10
bar*:
The
cylinder
pressures
are
shown
on the
screen.
2
Close
the
cylinder
vaives
again.
If
the
valves
remain
open
when
connected
to
the
central
gas
supply,
gas
may
be
withdrawn
from
the
backup
cylinders.
The
gas
supplies
available
can
be
selected
in
the
standby
configuration
(see
page
113).
Only
these
gas
supplies
will
then
be
checked
during
the
self
test
and
an
alarm
issued
in
the
event
of
a
fault
during
normal
operation.
The
central
oxygen
supply
and
the
O2
cylinder
cannot
both
be
configured
as
not
present
at
the
same
time.
Open
the
backup
gas
cylinders
which
have
been
configured
as
present
for
the
self
test
and
then
close
them.
O2
must
be
connected
for
the
following
self
test.
*
See
page
181
for
precise
details
on
cylinder
pressures.
Instructions
for
Use
Primus
SW
2.n
Getting
started
ikdger
Primus
|
U
see
000
y
)
ο ο
o
|
35
uni
0014
uni.0015

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup

Related product manuals