Dräger Primus User Manual

Dräger Primus
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
Getting
started
02
flush
1
Close
the
Y-piece
=
plug
firmly
onto
the
cone.
2
Press
the
»02
button.
3
Breathing
bag
inflates
with
an
audible
flow.
O2
emergency
delivery
1
Close
the
Y-piece
=
plug
firmly
onto
the
cone.
4
Press
the
»Safety O2«
knob
to
disengage
for
O2
emergency
delivery
and
turn
to
adjust
flow.
3
Breathing
bag
inflates
with
an
audible
flow.
4
Turn
»Safety
O2«
knob
back
to
its
original
position
to
discontinue
emergency
delivery
and
press
it
inwards.
Emergency
ventilating
bag
Example:
Dräger
Resutator
2000
@
Ready
for
operation.
Vaporisers
Note
the
corresponding,
individual
Instructions
for
Use
when
using
the
Vapor
19.3,
Vapor
2000
or
Devapor.
The
Vapor
2000
is
shown
and
described
here.
1
Locking
lever
points
to
the
left
=
locked.
©
Vapor
is
mounted
straight
and
seated
securly
on
the
vaporiser
connection.
2
Filling
level
is
adequate.
3
Handwheel
set
to
»0«
and
engaged.
4
Unused
vaporiser
is
locked
by
means
of
interlock
slide
(example:
left-hand
Vapor
locked).
After
filling
or
changing
the
vaporiser:
@
Perform
leak
test,
see
page
73.
Note
vaporiser
flow
limits!
Example
Vapor
2000:
0.25
to15
L/min
or
0.25
to10
L/min
at
a
concentration
>5
%
36
tit
LLELLL
ven
00.
Or
00
o
2
4
*f
©
©
ОГ]
instructions
for
Use
Primus
SW
2.п
uni_0017

Other manuals for Dräger Primus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Primus and is the answer not in the manual?

Dräger Primus Specifications

General IconGeneral
TypeAnesthesia Workstation
Oxygen FlushYes
Ventilation ModesPressure Support
Gas SupplyO2, Air, N2O
VaporizersIsoflurane, Sevoflurane, Halothane
Patient MonitoringAirway Pressure
Backup BatteryYes
AlarmsApnea, Gas supply failure
Power SupplyAC Mains, Battery Backup