EasyManua.ls Logo

Dräger REGARD-1 - Span Calibration with Calibration Gas; Gas Concentration to Voltage Mapping

Dräger REGARD-1
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration and calibration
Issue 5 - June 2005 Page 34 of 64 Pages
Connect a voltmeter across TP1 and TP2 on the main circuit board:
If applying calibration gas to the transmitter:
Apply calibration gas to the transmitter. Use a gas concentration that is between
40% and 90% of measuring range. The transmitter must be correctly calibrated.
Adjust potentiometer FS on the main circuit board until the voltage across TP1 and
TP2 corresponds to the gas concentration. If the controller has a display, confirm
that the reading is correct.
Gas concentration
(as % of measuring range)
Voltage across
TP1 & TP2 (V)
40 2.08
50 2.40
60 2.72
70 3.04
80 3.36
90 3.68

Other manuals for Dräger REGARD-1

Related product manuals