PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
5 - 3
High Temp alarm indicator When the skin probe is attached to the infant and the skin temper-
ature exceeds 39.0°C (102.2°F), the indicator flashes, the alarm
sounds continuously, and the heater turns off. To silence the alarm
for 2 min, press the Silence/Reset key. When the skin temperature
falls to 38.5°C (101.3°F), the alarm automatically resets.
Probe alarm indicator In Baby Mode, if the skin probe fails or is worn out, the indicator
flashes, and a three-level alarm sounds. After the skin probe is
replaced, the alarm automatically resets. If the skin probe fails due
to a shorted probe, the System Fail indicator lights, and an alarm
sounds. You cannot silence this alarm. To reset this alarm, switch
off the power, and then turn it back on.
Chk Patient alarm indicator In Manual Mode, if the warmer is in operation for longer than
10 min, the indicator illuminates and the alarm sounds one time.
Thereafter, the Chk Patient indicator remains on, and the alarm
sounds every 30 s for the next 5 min. To silence the Chk Patient
alarm for 10 min, press the Silence/Reset key within 5 min. If, at
the end of 5 min (15 min total), the Silence/Reset key has not
been pressed, the heater shuts down, and a three-level alarm
sounds continuously.
System Fail alarm indicator When an internal malfunction or operator error is detected, the
indicator illuminates, and the alarm sounds continuously. In addi-
tion, an error code, such as Er01 to Er023, appears in the Baby
Temperature display. When a malfunction is detected, the con-
troller automatically performs the selftest function to determine
whether the fault corrects itself. If the fault does not correct itself,
the error code appears until it is corrected. This alarm is not reset-
table. Refer the unit to qualified service personnel or correct oper-
ator error.
NOTE:
The operator can correct Error Code 023 by reducing the
ambient temperature to below 32 °C (90 °F).
Power Fail alarm indicator When power to the unit is interrupted while the controller is on,
the indicator flashes, and the alarm beeps. When power is restored
to the unit, the alarm automatically resets. Push the power key to
silence the alarm.
Procedural Silence indicator Indicates that the unit is in procedural silence. The procedural
silence interval is 5 min. During procedural silence, the low Baby
Temp alarms are blocked.