12
Operation
Operation
Switching on the device
Press and hold the key for approx. 3 seconds until the countdown » 3 . 2 . 1 « shown in
the display has elapsed.
All the display segments, including the visual, audible and vibration alarms, are activated for
a short time.
The software version is displayed.
The device performs a self test.
The remaining operating time (not in the case of X-am 2000) is displayed, e.g. » d 730 «
(remaining operating time 730 days).
All alarm setpoints A1 and A2 as well as TWA
*)
and STEL
*)
for H
2
S and CO are displayed
in succession.
The sensor which is next due for calibration/adjustment is displayed with the days remai-
ning until the next calibration/adjustment e.g. » Ex %UEG CAL 20 «.
During the warm-up time of the sensors, the respective display of the measured value
flashes and the special symbol » « (for warning) is displayed.
Press the key to cancel the display of the activation sequence.
Switching off the device
Press and hold the key and the key at the same time until the countdown » 3 . 2 . 1 «
shown in the display has elapsed.
Before the device is switched off, the visual, audible and vibration alarms are activated for
a short time.
*)
Only when activated in the device configuration. Delivery status: not activated.
OK
on
OK
on
OK
on
M
+

Table of Contents

Other manuals for Dräger X-am 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-am 2000 and is the answer not in the manual?

Dräger X-am 2000 Specifications

General IconGeneral
BrandDräger
ModelX-am 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals