Faults, Cause and Remedy
Fault messages
551 DrägerSensor XXS CO in the
warm-up phase
Wait until warm-up time is
complete.
552 DrägerSensor XXS CO in the
warm-up phase
Wait until warm-up time is
complete.
553 CO concentration has drifted
into the negative range
Carry out the fresh air
calibration, page 35.
554 The temperature is too high Operate the device within the
allowed temperature range.
555 The temperature is too low Operate the device within the
allowed temperature range.
556 The calibration interval for
DrägerSensor XXS CO has
elapsed
Carry out the span calibration
for DrägerSensor XXS CO,
page 39.
Special symbol » « and
displayed numerical code:
Cause Remedy
101 The service life of the device
has elapsed
Return the device to Dräger
Safety to be disposed.
102 The customer's service life
counter has elapsed
Reset the service life counter
using "CC Vision".
103 The device is defective The device must be repaired by
DrägerService.
104 Check sum error program
code
The device must be repaired by
DrägerService.
105 + 106 Check sum error data Reconfigure the device with "CC
Vision", otherwise return device
to DrägerService.
107 The device is defective The device must be repaired by
DrägerService.
108 The data base is defective Execute the menu function again;
otherwise the device must be
repaired by DrägerService.
109 The menu function cannot be
carried out because of an
error.
Determine the error code via the
info menu and switch it off, if
necessary.
Special symbol » « and
displayed numerical code:
Cause Remedy