Maintenance
Calibrating / Adjusting the Device
Allow the sensors to warm up before the calibration / adjustment!
Warm-up time: refer to the Instructions for Use of the installed DrägerSensors (on CD).
Calibration interval:
Observe the relevant specifications in the Instructions for Use of the installed
DrägerSensors.
For critical applications, observe the recommendations in EN 50073
*)
or EN 45544-4
**)
and national regulations.
Improve the zero point accuracy – carry out the fresh air calibration, page 35.
Set the sensitivity of all sensors to the value of the test gas – carry out the 1-button
calibration, page 37.
Set the sensor sensitivity to a value of the test gas – calibrate/adjust the sensitivity,
page 39.
*)
EN 50073 – Guidelines for selection, installation, use and maintenance of devices for the detection and
measurement of flammable gases and oxygen.
**)
EN 45544-4 – Electrical devices for the direct detection and direct concentration measurement of toxic
gases and vapours – Part 4: Guidelines for selection, installation, use and maintenance.
CAUTION:
Do not inhale the test gas. Observe the hazard warnings of the relevant Safety Data
Sheets as well as the Instructions for Use of the used DrägerSensors.

Table of Contents

Other manuals for Dräger X-am 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-am 2000 and is the answer not in the manual?

Dräger X-am 2000 Specifications

General IconGeneral
BrandDräger
ModelX-am 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals