Maintenance
When the calibration has been
completed:
— The display containing the current gas
concentration changes with the display
» OK «.
● Press the key to confirm the
calibration or wait for 5 seconds.
If a fault has occurred during the fresh air
calibration.
— The fault message » « appears and
» « is displayed for the respective
sensor instead of the measured value
● In this case, repeat the fresh air
calibration.
● If necessary, replace the sensor,
page 45.