EasyManuals Logo

Dräger X-am 2000 Technical Handbook

Dräger X-am 2000
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
40
Maintenance
Press the key to carry out the
calibration of the selected measuring
channel.
Enter the concentration of the test gas –
procedure similar to the enter password
procedure.
Complete the entry of the concentration
by pressing the key.
Open the test gas cylinder valve to let test
gas flow over the sensor.
The displayed, flashing measurement
value changes to the value according to
the supplied test gas.
When the displayed value has stabilized:
Press the key.
The display containing the current gas
concentration changes with the display
» OK «.
Press the key to exit the calibration.
The next measuring channel is offered for
calibration.
After the last measuring channel has
been calibrated/adjusted, the device
switches to measuring mode.
Close the test gas cylinder valve and
remove the device from the calibration
cradle.
If a fault has occurred during the span
calibration.
The fault message » « appears and
» « is displayed for the respective
sensor instead of the measured value
In this case, repeat the calibration.
If necessary, replace the sensor,
page 45.
OK
on
OK
on
03623854
_
01
_
en.e
p
s
OK
on
OK
on
03723854
_
01
_
en.e
p
s

Table of Contents

Other manuals for Dräger X-am 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-am 2000 and is the answer not in the manual?

Dräger X-am 2000 Specifications

General IconGeneral
BrandDräger
ModelX-am 2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals