23
DRAMIŃSKI iScan mini – user guide
11. System settings
11.1. Frame averaging
This function enables anti-aliasing by super-imposing every two or every four neighbouring frames and dis-
playing them simultaneously.
In order to change the frame averaging:
1. Enter the Menu and choose Settings.
2. Select Frame averaging.
3. Using the arrows, select a proper setting.
4. Conrm with the OK button.
11.2. LuciD
LuciD is a tool which causes: smoother structure edges, enhancement of parenchymal structures, improve-
ment of contrast.
In order to turn LuciD on/o:
1. Enter the Menu and choose LuciD.
2. Using the arrows, select a proper setting.
3. Conrm with the OK button.
11.3. Interleaving
The function enables choosing between an image composed of 128 lines (faster refresh rate, lower resolution)
and 256 lines (higher resolution, but lower refresh rate — default setting).
In order to change the interleaving setting:
1. Enter the Menu and choose Settings.
2. Select Interleaving.
3. Using the arrows, select a proper setting.
4. Conrm with the OK button.
11.4. Change left/right, up/down
This option enables setting the image so that the marker on the screen corresponds to the side of the marker
on the probe. It is meaningful for right- and left-handed people during rectal examination. The image may also
be reversed upside-down.
1. Enter the Menu and choose Settings.
2. Select L-R or U-D.
3. Using the arrows, select a proper setting.
4. Conrm with the OK button.
11.5. Rotate image by 90°
The option enables rotating the image by 90°. In order to use this function more comfortably, attach the carry-
ing belts so that the orientation of ultrasound scanner matches the orientation of the image.