DRAMIŃSKI iScan mini user guide
13. Maintenance of the device and the probes
13.1. Cleaning and disinfection
Attention! In view of the biological security, it is recommended to clean and disinfect the ultrasound
scanner after you nish working with it.
During use, the device may become strongly contaminated, including contagious material. Immediately after
nishing work, the scanner should be cleaned with a wet soft cloth or a paper towel with mild detergent.
The device surface should be disinfected with appropriate substance intended for disinfection of medical and
veterinary product surfaces.
Attention! The ultrasound probe should be carefully disinfected after every use.
After wet-cleaning, the probe and the ultrasound scanner should be wiped dry with a soft paper towel, if nec-
essary.
Warning! It is forbidden to use highly concentrated, aggressive and abrasive agents.
Such agents can permanently damage the surface of the probe, the display screen
and the casing.
During wet-cleaning, protect the casing sockets and ventilation slots against hu-
midity.
Warning! Use alcohol-free agents to disinfect the probe.
In order to wash the ultrasound scanner after work has been nished:
1. Protect the goggles and data transmission sockets with plugs.
2. Disconnect the battery.
3. Disconnect the carrying belts.
4. Wipe the ultrasound scanner with a damp (water or mild detergent) cloth or paper towel. Do not wash the
ultrasound scanner under running water!
5. Wipe the ultrasound scanner dry.
6. For disinfection, use a proper agent intended for disinfection of surfaces of medical and veterinary devices.
Follow the instructions on the label.
7. Clean and disinfect the battery and the probe using damp paper towels or tissues.
8. Let the ultrasound scanner, probe and battery dry in a safe place.
Attention! The transport case has a vent valve to control the air pressure in the closed case. It is not
moisture-permeable. Do not close a wet device in the case. If humidity accumulates in the case, we rec-
ommend using car desiccants available in stores and petrol stations.
13.2. Technical inspections
Remember to always check the ultrasound scanner, the battery, the probe and its cable for mechanical dam-
age before starting to work. If you notice anything disturbing (for example, discontinuity of the cable), contact
the service centre immediately.
The users of the ultrasound scanner are recommended to have regular technical inspections performed by the
manufacturer every two years. It will guarantee the highest safety and durability.