Dynatron® 850plus & 550plus
Treatment Instructions - MICROCURRENT
45
incremental value representing the volume selection. You must continue holding the
FUNCTION key down while adjusting the volume. When you release the FUNCTION
key, the Power-Intensity display returns to its normal display.
9. MODIFY a treatment in progress.
While the treatment is in progress, you can modify the treatment parameters.
Press the FUNCTION-MICRO key to select a different polarity.
Use the Time toggle key to select TIME then use the Time arrow keys to increase or
decrease the treatment time.
Use the Time toggle key to select FREQ, then use the Time arrow keys to increase or
decrease the frequency.
Use the Power-Intensity arrow keys to increase or decrease the intensity.
10. Press STOP, if it is necessary to end the treatment before the treatment time has expired.
When the treatment time has elapsed, current to the patient is stopped and a tone sounds
notifying you of the treatment end. Treatments in progress may be stopped at any time
using one of the following methods.
STOP: Press this key to stop treatment at all channels. You may prefer to stop treatment
at one channel only (see explanation below).
STOP ONE TREATMENT ONLY: If you have more than one treatment in progress,
you can stop one treatment by either of the following methods. First, press the channel
toggle key to select the channel to be stopped (that channel’s light is green when
selected).
FUNCTION-STOP. Press and hold the FUNCTION key and press STOP.
REDUCE THE TREATMENT TIME TO ZERO. Press the Time down arrow
until the Time display reaches zero. The device beeps when the time reaches
zero.
The output at the selected channel is stopped, and the device then displays the parameters for
the next treatment that remains in progress (if any).
Remote Stop Cable With Microcurrent
Even though microcurrent is a very low voltage therapy, you may allow the patient to hold the
Remote Stop cable during an unattended treatment if desired. If the button on the Remote Stop
cable is pressed during any microcurrent treatment, the treatment time and output are stopped, and
the timer returns to zero.