4.6 CONFIGURATION/CALIBRATION MODE
The configuration/calibration mode is enabled by opening the Mode Selection Switch on the left
side of the lower board (see Figure 7). A list of all configurable parameters starts below.
When the unit is turned on in configuration/calibration mode, the display will show
1290
and then
COnF
.
Press F to initiate the configuration procedure, starting at the first parameter.
Press R to initiate the configuration procedure, starting at the last parameter.
Press to toggle the display from
COnF
to
CAL
, initiating output calibration mode, if required.
Pressing will display the software version number.
For each parameter you will either select one of several choices or enter a numerical value.
Press to modify or change the parameter or increase the number displayed.
Press to modify or change the parameter or decrease the number displayed.
Press F to save your changes and step to the next parameter.
Press R to step to the previous parameter without saving your changes.
For many parameter settings, the display initially alternates between a code and a numerical value.
Once you start to modify the value, however, only the numerical value will be shown.
When entering a numerical value, and are used to increase or decrease the number being
entered. The change in value for each step is either 1, 10, or 100, depending on the size of the
number. The increment is 1 for values up to 2010, 10 from 2010 to 20100, 100 from 20100 to
99900.
4.7 CONFIGURABLE PARAMETERS
The following is the complete sequence of configurable parameters. Default values are given on
page 24.
A. LINE FREQUENCY
The display shows
L.F.
(Line Frequency) followed by:
60
for 60 Hz, or
50
for 50 Hz
B. DECIMAL POINT POSITION
The display shows:
-----
for no digits after decimal point
----.-
for one digit after decimal point
---.--
for two digits after decimal point
--.---
for three digits after decimal point
-.----
for four digits after decimal point