26
Er 39
Error detected during EAROM write operation. This error may appear in operating mode when
storing a new value in EAROM (for example, alarm threshold or transducer calibration). the new
values will be enabled but they will be lost when the instrument is powered down. This error
message disappears automatically after 10 seconds.
If this error appears during configuration/calibration, press F or R to restart the procedure and then
repeat operations. If the error persists, return the instrument to your supplier.
Er 101
The configuration data stored in EAROM is wrong or inconsistent. The error may appear at
instrument startup in operating mode. This error message disappears automatically after 3 seconds,
and the instrument resets.
If the error persists, enable configuration/calibration mode with the internal switch, load the default
calibration data and then perform a new configuration.
Er 201
The calibration data stored in EAROM for analog retransmission is wrong. The error may appear at
instrument startup in operating mode. This error message disappears automatically after 3 seconds,
and the instrument resets.
If the error persists, enable configuration/calibration mode with the internal switch, load the default
calibration data and then perform a new analog retransmission calibration.
Er 312
Error during internal autozero measurement for temperature drift compensation. The instrument
repeats this check every 3 seconds. The analog retransmission and alarm go low scale or high scale
as a failsafe configuration. If the error persists, return the instrument to your supplier.
Er 402
The configuration/calibration data stored in EAROM is not protected. The error may appear at
instrument startup in operating mode. This error message disappears automatically after 3 seconds,
and the instrument resets.
If the error persists, enable configuration/calibration mode with the internal switch, power up the
unit and then return to operating mode. This action should be sufficient to enable data protection.
o o o o o
Over-range indication