3. SCALING
Insert instrument in case and tighten screw.
Display will indicate 1290 then CONF.
Push F key three times. Display will read F.S.U. (Full Scale Value).
Using the Down arrow key, select the value that matches the full scale of the transducer. Enter
with F key. (If your transducer is 10,000 PSI, no adjustment is necessary.)
Remove instrument from case and close mode selection switch as shown.
Reinsert instrument into case and tighten screw.
4. CALIBRATION AND OPERATION
•When power is applied, display will read 1290 then a Pressure reading. At this point, be sure
that no pressure is applied and that the transducer is at operating temperature.
Press the F key and the legend LO. OFF appears. Using the Up arrow key, advance the legend to
LO.ON.
Press the F key to initiate zero calibration. After approximately 5 seconds, the legend F.U.S.C.
will appear and alternate with the full-scale value.
Press the Up arrow key to advance to the legend FU ON. using the F key, initiate the full-scale
calibration. After 5 seconds, the indicator will read 0 ±10.
•The system is now ready for use.