User manual SEFELEC 5x Series  182   v1.06 
 
OHM (OHMmeter) :  
Sets the main thresholds unit (Ohm) and the ground continuity display function.  
Caution : The action on this command will erase all previous recorded 
thresholds (Reset to zero). 
 
VOLT (VOLTmeter) :  
Sets the main thresholds unit (Volt) and the ground continuity display function.  
Caution : The action on this command will erase all previous recorded 
thresholds (Reset to zero). 
 
HTIM (HTIMe) <NR2> :  
Sets a new value in second to the current hold time parameter to the parameter set 
of the running function.  
For instance : HTIM 3.2 
 
RTIM (RTIMe) <NR2> :  
Sets a new value in second to the current rise time parameter to the parameter set 
of the running function 
For instance : RTIM 10.5 
 
FTIM (FTIMe) <NR2> :  
Sets a new value to the current fall time parameter to the parameter set of the run-
ning function. 
For instance : FTIM 0.5 
 
HLIM (HLIMit) <NR3> :  
Sets a new value to the high limit parameter to the parameter set of the running 
function. 
For instance :   
  MEG:HLIM 2.0E+6  (new value in Ohms) 
 HIP:HLIM 1.45E-4   (new value in Amps) 
 GND:HLIM 1.00E-1     (new value in milliohm) 
The unity is automatic, given in Volt, Ohm or Amp and function dependent. 
 
CB HLIM <NR3> :  
Same command as HLIM, but reset the low limit parameter before to change the 
high limit value. 
 
LLIM (LLIMit) <NR3> : 
Sets a new value to the low limit parameter to the parameter set of the running 
function. 
For instance :   
  MEG:LLIM 1.0E+6     (new value in Ohms) 
 HIP:LLIM 3.50E-6      (new value in Amps) 
 GND:LLIM 5.02E-2     (new value in milliohm) 
The unity is automatic, given in Volt, Ohm or Amp and function dependent. 
 
TIM (TIMe) AUT/PERM  
Select the temporization mode. AUT for automatic, PERM for Permanent test until 
a fault occurs.