User manual SEFELEC 5x Series  181   v1.06 
 
18.1.4.  COMMANDS DEPENDENT ON THE DEVICE 
MEG (MEGohmmeter) :  
Selects the megohmmeter. Must be sent from the initialization display. 
 
HIP (HIPot) :  
Selects the Hipot function. Must be sent from the initialization display. 
 
GND (GrouND) :  
Selects ground continuity function. Must be sent from the initialization display. 
 
SEQ (SEQuence) :  
Selects the sequence function. Must be sent from the initialization display. 
 
CONF (CONFig) :  
Selects the configuration function of the unit. Must be sent from the initialization 
display. 
 
PAR (PARameter) <NR1> :  
Selects the parameter set for the running function. 
i.e.: MEG : PAR 1 
 
DCV (DCVoltage) <NR1> :  
Sets a new value in volt to the current DC voltage parameter to the parameter set 
of the running function. 
i.e. : DCV 500 
 
ACV (ACVoltage) <NR1> : 
Sets a new value in volt to the current AC voltage parameter to the parameter set 
of the running function without changing the pre-set frequency. 
Hipot tester : Min. : 100 Max. : 5000  
For instance : hipot : ACV 5000     
 
ACV50 (ACVoltage) <NR1> : 
Sets a new value in volt to the current AC (50Hz frequency) voltage parameter to 
the parameter set of the running function 
Hipot tester : Min. : 100 Max. : 5000  
For instance : hipot : ACV50 5000 
 
ACV60 (ACVoltage) <NR1> : 
Sets a new value in volt to the current AC (60Hz frequency) voltage parameter to 
the parameter set of the running function 
Hipot tester: Min.: 100 Max. : 5000  
For instance: hipot : ACV60 5000 
 
ACC (ACCurrent) <NR3> :  
Sets a new value in ampere to the AC current parameter to the parameter set of 
the ground continuity function.