APPENDIX A: EMC INFORMATION
If you have any questions regarding the guidance and declarations listed below, please contact your home
care provider.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions: This device is intended for use in
the electromagnetic environment specifi ed below. The user of this device should make sure that it is used
in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC
ENVIRONMENTGUIDANCE
RF emissions CISPR 11 Group 1 This device uses RF energy only
for its internal function. Therefo-
re, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B The device is suitable for use
in all establishments, including
domestic establishments and
those directly connected to the
public low-voltage power supply
network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fl uctuations/Flicker Complies
emissions IEC 61000-3-3
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: This device is intended for use in
the electromagnetic environment specifi ed below. The user of this device should make sure that it is used
in such an environment.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC
ENVIRONMENT
GUIDANCE
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±15kV air
±8kV contact
±15kV air
±8kV contact
Floors should be wood,
concrete or ceramic tile.
If fl oors are covered with
synthetic material, the rela-
tive humidity should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2kV for power
supply lines
±1kV for input-output
lines
±2 kV for supply
mains
±1 kV for input/output
lines
Mains power quality
should be that of a ty-
pical home or hospital
environment.
Surge
IEC 61000-4-5
±1 kV di erential
mode
±2 kV common
mode
±1 kV di erential
mode
±2 kV common
mode
Mains power quality
should be that of a ty-
pical home or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95% dip in
UT) for 0.5 cycle 40%
UT (60% dip in UT) for 5
cycles 70% UT
(30% dip in UT) for 25
cycles <5% UT
(>95% dip in UT) for 5
seconds
<5% UT (>95% dip in
UT) for 0.5 cycle 40%
UT (60% dip in UT)
for5 cycles 70% UT
(30% dip in UT) for 25
cycles <5% UT
(>95% dip in UT) for 5
seconds
Mains power quality
should be that of a ty-
pical home or hospital
environment.
If the user of the de-
vice requires continued
operation during power
mains interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or battery.
Power frequency
(50/60 Hz) magnetic
fi eld IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic
fi elds should be at levels
characteristic of a typical
location in a typical hospi-
tal or home environment.