EasyManua.ls Logo

Ecom j2kn pro - Page 43

Ecom j2kn pro
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ecom-J2KN
pro
Page 43
Press Start measurement“ / <OK> to
start the real measurement. Proceed as
follows:
- select surface (boiler side) to be measured
with <F1> or <F2>
- position surface sensor
- record temperature difference with
<Memory> - up to 10 values can be
recorded per surface out of which a mean
value will be calculated automatically
- if need be, cancel measurements
with <F4>
- repeat this procedure for each surface
Once all surface temperatures have been determined, quit the menu
with <ESC>. The surface losses get automatically calculated. The
value conversion in negative points is available in the menu point
„Results“.
The ventilation loss measurement is
performed by a flow probe 30 sec. after
burner switch-off. This measurement
can be performed at the earliest 5 min.
after instrument´s switch-on as the
pressure sensor requires this period of
time for stabilisation. Once the menu is
called up, the values for air pressure,
external temperature, boiler perfor-
mance and exhaust gas pipe diameter
must be entered. Hereto proceed as
follows:
- open respective window with <OK>
- input figures using numerical function
of keys
- confirm input with <OK>
Width --- m
Height --- m
Surf --- m2
T.S. --- °C
T.A. --- °C
T.S. 21.5 °C
T.A. 21.5 °C
Surface losses
-
-
Flow probe
Width 1.20 m
Height 1.20 m
Surf 1.44 m2
T.S. 40.5 °C
T.A. 21.5 °C
T.S. 42.5 °C
T.A. 21.5 °C
Surface losses
-
-

Other manuals for Ecom j2kn pro

Related product manuals