SE-2003&SE-2012 Series Holter System Recorder User Manual
- 37 -
Electromagnetic Immunity
Guidance and manufacture's declaration electromagnetic immunity
Holter System Recorder is intended for use in the electromagnetic environment specified below.
The customer or the user of Holter System Recorder should assure that it is used in such an
environment.
Immunity test
IEC/EN 60601
test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
±8 kV contact
±15 kV air
±8 kV
contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC/EN 61000-4-4
2 kV for power
supply lines
1 kV for
input/output
lines
Not
applicable
Not applicable
Surge
IEC/EN 61000-4-5
1 kV line to line
2 kV line to
ground
Not
applicable
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC/EN 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.