Patient Monitor User Manual Product Specification
- 252 -
Adult Default: 160 mmHg
Range: 80/100/120/140/150/160/180/200/220/240 mmHg
Pediatric Default: 140 mmHg
Range: 80/100/120/140/150/160/180/200 mmHg
Neonatal Default: 100 mmHg
Range: 60/70/80/100/120 mmHg
Venipuncture pressure
Adult Default: 60 mmHg
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50 mmHg, 60
mmHg, 70 mmHg, 80 mmHg, 90 mmHg, 100 mmHg, 110
mmHg, 120 mmHg
Pediatric Default: 40 mmHg
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50 mmHg, 60
mmHg, 70 mmHg, 80 mmHg
Neonatal Default: 30 mmHg
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50 mmHg
Omron Module
Not applicable to iM50, iM60 or M50.
Method Oscillometric
Mode Manual, Auto, Continuous, Sequence
Measuring Interval in AUTO Mode
(unit: minute)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480 and
User Define
Continuous 5 min, interval is 5 s
☆Measuring Parameter
SYS, DIA, MAP
Pressure Unit kPa, mmHg, cmH
2
O
☆Measuring Range
☆Adult/ Pediatric Mode
SYS: 60 mmHg to 250 mmHg
DIA: 40 mmHg to 200 mmHg
MAP: 45 mmHg to 235 mmHg
☆Neonatal Mode
SYS: 40 mmHg to 120 mmHg
DIA: 20 mmHg to 90 mmHg
MAP: 30 mmHg to 100 mmHg